SLE 5000 artificial lung ventilation apparatus for newborns
The SLE 5000 artificial lung ventilation apparatus is a high-end device, the result of the constant search for new solutions to achieve an effective combination of quality care for the patient and spent money on it. It is designed to replace and support the function of breathing in newborns, including those with ELBW and children of early age (from 300 g to 20 kg)
Advantages
Ventilation Modes
Specifications
Traditional ventilation modes CPAP / PTV / PSV | |
Inhalation time | 0.1 - 3.0 sec. |
CPAP pressure | 0 - 20 mbar |
Inhalation pressure | 0 - 65 mbar |
Specified volume | 2 - 200 ml |
FiO2 | 21% - 100% |
CMV / SIMV | |
BPM (breathing rhythm) | 1-150 inhalation/min. |
Inhalation/expiration rate (I:E) | 11.2:1 - 1:600 |
Inhalation time | 0.1 - 3.0 sec. |
PEEP pressure | 0 - 20 mbar |
Inhalation pressure | 0 - 65 mbar |
Specified volume | 2 - 200 ml |
FiO2 | 21% - 100% |
Ventilation Modes: HFO High-frequency ventilation | |
Frequency range | 3 - 20 Hz |
Inhalation/expiration rate (I:E) | 1:1 |
Delta Pressure Range | 4 - 180 mbar |
Mean pressure range | 0 - 35 mbar |
FiO2 | 21% - 100% |
HFO+CMV | |
Breathing rate BPM | 1 - 150 bpm |
Inhalation time | 0.1 - 3.0 sec. |
Frequency range | 3 - 20 Hz |
Inhalation/expiration rate | 1:1 / 1:2 / 1:3 |
Inhalation pressure | 0 - 65 mbar |
Delta Pressure Range | 4 - 160 mbar |
Mean pressure range | 4 - 160 mbar |
FiO2 | 21% - 100% |
Monitoring Parameters of Flow and Volume Measurement | |
Flow sensor type - double heated anemometer (autoclavable or disposable) | 10 mm |
Flow (Accuracy±8%) | 0.2 - 32 l / min |
Expiratory tidal volume | 0 - 999 ml |
Expiratory minute volume | 0 - 18 l |
Dead spot | 1 ml |
Weight | 10 g |
Traditional Ventilation and Combined Modes: | |
Leakage in a patient | 0 - 50% (Resolution - 5%, average for the last 5 breaths) |
Total breathing rate | 0 - 150 inhalation/min. |
Dynamic compliance | 0 - 100 ml/mbar (Resolution - 1 ml/mbar) |
C20/C | resolution 0.1 |
Measurement time | 2 msec |
Resistance | 0-1000 mbar.sec / l |
Flow trigger sensitivity | 0.2 - 10 l / min |
Oxygen concentration
Range | 21 - 100% (resolution 1%) |
Pressure
Pressure measurement in real time | Resolution 1 mbar |
Measurement time | 2 msec |
Peak pressure | 0 - 175 mbar (Resolution - 1 mbar) |
PEEP pressure | 0 - 175 mbar (Resolution - 1 mbar) |
Mean pressure | -175 - 175 mbar (Resolution - 1 mbar) |
User-Defined Alarms
Peak pressure Auto install, when the patient pressure control is adjustable or can be adjusted manually | |
Range | 10 - 110 mbar |
Resolution | 0.5 mbar |
Cycle Failure Auto install, when the patient pressure control is adjustable or can be adjusted manually | |
Low pressure Auto install, when the patient pressure control is adjustable or can be adjusted manually | |
Range | 10 - 70 mbar |
Traditional modes, HFO modes | up to 10 mbar below the high pressure threshold |
Low Tidal Volume | |
Range | 0 - 200 ml |
Resolution | 0.2 ml |
Low Minute Volume | |
Range | from 0 to 0.02 liters below the High Minute Volume Threshold |
Resolution | 0.2 ml |
High Minute Volume | |
Range | 0.02 from 18 l |
Resolution | 0.1 l |
Apnea Time Set in CPAP or when the respiratory support rate is less than 20 BPM | |
Range | from 3 to 60 sec |
Resolution | 1 sec |
Power
Voltage | 100-250 V / 50-60 Hz |
Power | 115 VA |
Battery | 45-60 minutes |
Battery charge | Full charge for 24 hours, 80% charge - for 8 hours |
Outputs - RS-232C
Air and O2 connection
Pressure | 3 - 5 bar |
Fresh gas flow | 8 l/min |
Maximum gas flow | 60 l/min |
Application environment
Temperature of | 10 - 40 ºC |
Humidity | 0-90% (without condensation) |
Dimensions
Size, fan only | 330mm (L) x 330mm (H) x 470mm (D) |
Height on a short rack | 114 cm |
Height on a high rack | 131 cm |
Weight, fan without rack | 21.8 kg |